Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8600A: (1 year) CAL VER /5100,5200,5205,8842 DATE: 01-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 2.6 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 44 NUMBER OF LINES: 194 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 8842A CONFIGURATION: Fluke 5200A CONFIGURATION: Fluke 5205A ============================================================================= # # Source: # Fluke 8600A Manual # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # The default 5200A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # The 90 day specifications of the 5200A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Before performing verification allow the 8600A to 1.005 DISP Stabilize (30 min.) at 23+/-5degC. 1.005 DISP Humidity less than 80% 1.006 DISP Turn the UUT on. 1.007 ASK- N 1.008 HEAD {RESISTANCE TESTS} 1.009 HEAD RESISTANCE TESTS: {200 Ohm Range} 1.010 RNG kZ 1.011 DISP Short the V-Ohm HI and V-Ohm LO terminals. 1.012 MATH MEM1 = 0.0 1.013 MEMI Enter UUT reading in Ohms: 1.014 MEME 1.015 MEMC 200 Z +0.03U 2.001 DISP Remove the short. 2.002 HEAD RESISTANCE TESTS (Test Lead Compensation) 2.003 DISP Configure the 8842A as follows: 2.003 DISP [32] 8842A INPUT HI to 8842A SENSE HI 2.003 DISP [32] 8842A INPUT LO to 8842A SENSE LO 2.004 IEEE [@8842]G5[I] 2.005 MATH MEM1 = MEM - 1100 2.006 JMPF 2.009 2.007 DISP Select FRONT INPUTS on the 8842A. 2.008 JMP 2.004 2.009 DISP Connect the 5100B OUTPUT to the 8842A INPUT. 2.010 RESF 100Z S 2W 2.011 8842 100Z N 4W 2.012 MATH M[1] = MEM 2.013 RESF 1kZ S 2W 2.014 8842 1kZ N 4W 2.015 MATH M[2] = MEM 2.016 RESF 10kZ S 2W 2.017 8842 10kZ N 4W 2.018 MATH M[3] = MEM 2.019 RESF * S 2.020 HEAD RESISTANCE TESTS: 200 Ohm Range 2.021 DISP Disconnect the leads from the 8842A and connect them to 2.021 DISP the UUT. 2.022 RESF 100.0Z S 2W 2.023 ASK+ N 2.024 MATH MEM = M[1] # 8842A 90 day accuracy, 200 Ohm Range 2.025 ACC 100 Z 0.007% 0.004U 2.026 MEMI Enter UUT reading in Ohms: 2.027 MEME 2.028 MEMC 200 Z 0.13U 3.001 ASK- U 3.002 HEAD RESISTANCE TESTS: {2 kOhm Range} 3.003 RESF 1.000kZ S 2W 3.004 ASK+ U 3.005 MATH MEM = M[2] # 8842A 90 day accuracy, 2 kOhm Range 3.006 ACC 1 kZ .005% .00003U 3.007 MEMI Enter UUT reading in kilohms: 3.008 MEME 3.009 MEMC 2 kZ 0.0011U 4.001 ASK- U 4.002 HEAD RESISTANCE TESTS: {40 kOhm Range} 4.003 RESF 10.00kZ S 2W 4.004 ASK+ U 4.005 MATH MEM = M[3] # 8842A 90 day accuracy, 20 kOhm Range 4.006 ACC 10 kZ .005% .0003U 4.007 MEMI Enter UUT reading in kilohms: 4.008 MEME 4.009 MEMC 20 kZ 0.006U 5.001 ASK- R 5.002 HEAD RESISTANCE TESTS: {200 kOhm Range} 5.003 RESF 200 100.0kZ 0.06U 2W 6.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 6.002 RESF 2000 1000kZ 0.6U 2W 7.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 7.002 RESF 20 10.00MZ 0.021U 2W 8.001 ASK+ R 8.002 EVAL Is the display flashing and the 20 MOhm LED ON? 9.001 HEAD {DC VOLTAGE TESTS} 9.002 HEAD DC VOLTAGE TESTS: {200mV Range} 9.003 5100 200 190.00mV 0.1U 2W 10.001 5100 200 -19.00mV 0.03U 2W 11.001 HEAD DC VOLTAGE TESTS: {2V Range} 11.002 5100 2 1.9000V 0.0005U 2W 12.001 5100 2 -1.9000V 0.0005U 2W 13.001 5100 2 0.9000V 0.0003U 2W 14.001 HEAD DC VOLTAGE TESTS: {20V Range} 14.002 5100 20 19.000V 0.005U 2W 15.001 5100 20 -19.000V 0.005U 2W 16.001 HEAD DC VOLTAGE TESTS: {200V Range} 16.002 5100 200 190.00V 0.05U 2W 17.001 HEAD DC VOLTAGE TESTS: {1200V Range} 17.002 5100 1200 1000.0V 0.3U 2W 18.001 HEAD DC VOLTAGE TESTS: {Auto Range} 18.002 RNG V 18.003 ASK- R 18.004 5100 200 -170.00mV 0.09U 2W 19.001 5100 2 -1.7000V 0.0004U 2W 20.001 5100 20 -17.000V 0.004U 2W 21.001 5100 200 -170.00V 0.04U 2W 22.001 5100 1200 -1000.0V 0.3U 2W 23.001 ASK+ R 23.002 DISP Disconnect the 5100B from the UUT. 23.003 HEAD {AC VOLTAGE TESTS} 23.004 DISP Connect the 5200A and UUT as follows: 23.004 DISP [32] 5200A OUTPUT HI to UUT V-Ohm HI 23.004 DISP [32] 5200A OUTPUT LO to UUT V-Ohm LO 23.005 HEAD AC VOLTAGE TESTS: {200mV Range} 23.006 5200 200 190.00mV 0.54U 500H 2W 24.001 5200 200 39.00mV 0.4U 30kH 2W 25.001 5200 200 190.00mV 1.95U 70kH 2W 26.001 HEAD AC VOLTAGE TESTS: {2V Range} 26.002 5200 2 1.0000V 0.0055U 50H 2W 27.001 5200 2 0.9000V 0.0021U 500H 2W 28.001 5200 2 1.9000V 0.0041U 500H 2W 29.001 5200 2 1.9000V 0.010U 50kH 2W 30.001 5200 2 1.9000V 0.02U 100kH 2W 31.001 HEAD AC VOLTAGE TESTS: {20V Range} 31.002 5200 20 19.000V 0.041U 500H 2W 32.001 5200 20 19.000V 0.2U 70kH 2W 33.001 HEAD AC VOLTAGE TESTS: {200V Range} 33.002 5200 200 100.00V 0.55U 30kH 2W 34.001 DISP Disconnect the 5200A from the UUT. 34.002 DISP Connect the 5205A and UUT as follows: 34.002 DISP [32] 5205A OUTPUT HI to UUT V-Ohm HI 34.002 DISP [32] 5205A OUTPUT LO to UUT V-Ohm LO 34.003 5200 200 190.00V 0.41U 500H 2W 35.001 HEAD AC VOLTAGE TESTS: {1200V Range} 35.002 5200 1200 1000.0V 3.1U 500H 2W 36.001 5200 1200 1000.0V 6.0U 20kH 2W 37.001 DISP Disconnect the 5205A from the UUT. 37.002 HEAD {DC CURRENT TESTS} 37.003 DISP Connect the 5100B and UUT as follows: 37.003 DISP [32] 5100B OUTPUT HI to UUT mA HI 37.003 DISP [32] 5100B OUTPUT LO to UUT mA LO 37.004 HEAD DC CURRENT TESTS: {200uA Range} 37.005 RNG 200uA 37.006 MATH MEM1 = 0.0 37.007 MEMI Enter UUT reading in microamps: 37.008 MEME 37.009 MEMC 200 uA 0.05U 38.001 5100 200 190.00uA 0.21U 2W 39.001 5100 200 190.00uA 0.21U 2W 40.001 HEAD DC CURRENT TESTS: {2mA Range} 40.002 5100 2 1.9000mA 0.0021U 2W 41.001 HEAD DC CURRENT TESTS: {20mA Range} 41.002 5100 20 19.000mA 0.021U 2W 42.001 HEAD DC CURRENT TESTS: {200mA Range} 42.002 5100 200 190.00mA 0.21U 2W 43.001 HEAD DC CURRENT TESTS: {2000mA Range} 43.002 5100 2000 1900.0mA 2.1U 2W 44.001 END